Name: | HYME-S hydrogen gas analyzer |
---|---|
Measuring range: | 0~1000ppm, 0~2%, 0~100% (optional) |
Resolution: | 1ppm, 0.01%, 0.1% |
Accuracy: | ≤±2%F.S. |
Signal output: | 4~20mA or 0~5V |
Communication interface: | Rs-232 or Rs-485 Modbus RTU |
Power supply: | 100~240VAC 50/60Hz |
Measuring range: 0~1000ppm, 0~2%, 0~100% (optional)
Resolution: 1ppm, 0.01%, 0.1% (according to the measuring range selected)
Accuracy: ≤±2%F.S.
Repeatability: ≤±1%F.S.
Stability:
Zero drift: ±1%F.S./7d
Span drift: ±1%F.S./7d
Response time T90≤30s
Warm-up time: ≤45min
Sample gas flow rate: 400 ± 50mL
Sample gas pressure: 0.1Mpa≤entrance pressure≤0.2Mpa
Power supply: 100~240VAC 50/60Hz 1A
Ambient temperature: -5℃~+45℃
Environment humidity: ≤90%RH
Output signal: 4~20mA or 0~5VDC
Communication interface: RS232 or RS485 (communication protocol: Modbus RTU)
Overall dimensions:
standard 160 (W) x160 (H) x280 (D) (mm)
Hole dimensions:
standard 138 (W) x138 (mm)
Weight: 2.5Kg
120*64 LCD monitors, desktop or panel mounted, easy to install
Imported TCD or electrochemical sensor, long service life, high sensitivity and fast response
Built in precision filter, greatly improving the sensor life
The upper and lower limit control points can be set arbitrarily within the full range
Data backup, data recovery function
Standard 4-20mA or 0-5V output
Standard RS232 / RS485 communication
All English operating menu
Built-in precision pump (optional)
HYME-S hydrogen gas analyzer adopts TCD or electrochemical gas sensor, combined with the monolithic integrated circuit technology, to analyze hydrogen gas content in mixture gases.
Petroleum, chemical, metallurgy, electric power, bio-fermentation and air separation industries
Hydrogen content non-explosion-proof online analysis in ammonia decomposition making hydrogen gas industry
Fertilizer analysis of hydrogen content in synthetic ammonia processes
Monitoring of the concentration of hydrogen in hydrogen cooling system in thermal power plant
Hydrogen leak in the containment of nuclear power plant monitoring and monitoring of the concentration of hydrogen in hydrogen-cooled system
Laboratory combustion testing of gas content determination
Gas stations or gas analysis of hydrogen purity in mixture gases